New TopShim run:

version = topshim 4.4.0
date = 2026-03-31
time = 09:31:42.363
command = topshim tunebxyz tuneaz ordmax=6

FINE SHIMMING

Parameters:

probe = Z113652_0123
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.5 Hz, Target= 0.2 Hz, After= 0.3 Hz
Iteration 2: FWHM Before= 0.3 Hz, Target= 0.6 Hz, After= 0.4 Hz
Iteration 3: FWHM Before= 0.4 Hz, Target= 0.8 Hz, After= 0.6 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.996


Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	-75
Z2	+54
Z3	-16
Z4	-100
Z5	-89
Z6	+109

duration = 1 min 27 sec
completed successfully
finished Tue Mar 31 09:33:09 2026


========================================================================