New TopShim run:

version = topshim 4.4.0
date = 2025-07-12
time = 06:57:20.786
command = topshim tunebxyz tuneaz ordmax=6

FINE SHIMMING

Parameters:

probe = Z113652_0123
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 6.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration 2: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration 3: FWHM Before= 0.4 Hz, Target= 0.4 Hz, After= 0.3 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.487


Performing tune after shimming
tune lock level increase = 1.1

shim changes: 
Z	-551
Z2	-174
Z3	-201
Z4	-1127
Z5	-843
Z6	-579

duration = 1 min 35 sec
completed successfully
finished Sat Jul 12 06:58:55 2025


========================================================================